High-precision CO₂ isotopologue spectrometer with a difference-frequency-generation laser source
A precision laser spectrometer for the detection of CO₂ isotopes is reported. The spectrometer measures the fundamental absorption signatures of ¹³C and ¹²C isotopes in CO₂ at 4.32 μm using a tunable mid-IR laser source based on difference-frequency generation. The spectrometer attains a precision of up to 0.02‰ for 150 s of averaging. An overall accuracy of 0.05‰ was obtained when sampling various calibrated reference gases.
document
http://n2t.net/ark:/85065/d7dn463r
eng
geoscientificInformation
Text
publication
2016-01-01T00:00:00Z
publication
2009-01-15T00:00:00Z
Copyright 2009 Optical Society of America. This paper was published in Optics Letters and is made available as an electronic reprint with the permission of OSA. The paper can be found at the following URL on the OSA website: http://www.opticsinfobase.org/ol/abstract.cfm?URI=ol-34-2-172. Systematic or multiple reproduction or distribution to multiple locations via electronic or other means is prohibited and is subject to penalties under law.